TC 6622 Front view
TC 6622 Front view
TC 6621 / 6622 (3) - AOIP, Instrumentation de test et mesure, contrôle moteur
TC 6621 / 6622 (4) - AOIP, Instrumentation de test et mesure, contrôle moteur
TC 6621 / 6622 (5) - AOIP, Instrumentation de test et mesure, contrôle moteur
Graph screen
Ramp simulation

TC 6621 / 6622

picto fleche Handheld temperature calibrator for thermocouples or resistive probes with memory

TC 6621 temperature calibrator measures or generates thermocouple temperatures (16 different types) and voltage at an accuracy of 0.02%.

TC 6622 temperature calibrator measures or generates RTD temperatures (12 different types) and resistance at an accuracy of 0.02%.

Ask a question
Get a call

Easy-to-use and equipped with a large graphical display, TC temperature calibrators are the perfect field tools for easy and quick maintenance and commissioning of temperature transmitters and probes.

TC 6621 temperature calibrator measures or generates thermocouple temperatures (16 different types) and voltage at an accuracy of 0.02%.

TC 6622 temperature calibrator measures or generates RTD temperatures (12 different types) and resistance at an accuracy of 0.02%.

With a very low temperature coefficient (10 ppm/°C in thermocouples and 7ppm/°C in resistance), IP54 protection and robust design, TC calibrators are suitable for onsite use even in demanding environmental conditions. They are widely used in the energy field, engineering sector, metal industry and automotive field.

TC calibrators use a graphical interface making programming and reading easier, under graphical or test format. Due to memory feature (10,000 values) and extended functionalities (square root, steps, synthesizer, statistical functions…), TC are well adapted to different process job procedures and ensure full data traceability as well as advanced data exploitation. Use them together with DATACAL calibration software to display, manage data and issue your own customized calibration certificates.

TC calibrators are delivered with 4 alkaline AA batteries in standard. The optional battery charger used a rechargeable battery.

Key features:

  • High accuracy: 0.02% reading with an adjustable resolution of 1 µV (TC 6621) and 1 mΩ (TC 6622)
  • Values displayed in °C, °F, mV and Ω
  • Low temperature coefficient: 10 ppm /°C for thermocouples and 7 ppm/°C for resistance
  • Measurements with HOLD function
  • Simulation of ramps, preprogrammed steps and synthetizer values
  • Correction of calibrated sensors
  • Display of minimum, maximum and average value
  • Backlight
  • 10,000 values stored and displayed graphically

Specifications and performances of TC 6621 @23°C ±5°C

DC voltage

Function Range Res. Accuracy / 1 year Measuring range
IN 100 mV 1 µV 0.020% RDG + 3 µV -10 mV / 100 mV
OUT 80 mV 1 µV 0.020% RDG + 3 µV -9.5 mV / 80 mV

Temperature coefficient < 15 ppm/°C beyond reference domain

Thermocouples: Measurement and simulation

Type Input range Resolution Accuracy / 1 year (Measurement) Output range Resolution Accuracy / 1 year (Simulation)
K -250 to -200°C
-200 to -120°C
-120 to 0°C
0 to +1372°C
0.1°C
0.1°C
0.05°C
0.05°C
0.90°C
0.3°C
0.02% RDG + 0.12°C
0.02% RDG + 0.11°C
-240 to -50°C
-50 to +120°C
+120 to +1372°C
0.1°C
0.1°C
0.05°C
0.8°C
0.3°C
0.02% RDG + 0.11°C
T -250 to -200°C
-200 to -50°C
-50 to +400°C
0.1°C
0.05°C
0.05°C
0.8°C
0.25°C
0.02% RDG + 0.09°C
-240 to -100°C
-100 to -40°C
-40 to +400°C
0.1°C
0.05°C
0.05°C
0.5°C
0.25°C
0.02% RDG + 0.10°C
J -210 to -200°C
-200 to -120°C
-120 to +60°C
+60 to +1200°C
0.05°C
0.05°C
0.05°C
0.05°C
0.3°C
0.25°C
0.02% RDG + 0.11°C
0.02% RDG + 0.09°C
-210 to +50°C
+50 to +500°C
+500 to +1200°C
0.05°C
0.05°C
0.05°C
0.35°C
0.02% RDG + 0.11°C
0.02% RDG + 0.09°C
E -250 to -200°C
-200 to -100°C
-100 to +450°C
+450 to 1000°C
0.1°C
0.05°C
0.05°C
0.05°C
0.55°C
0.2°C
0.02% RDG + 0.07°C
0.02% RDG + 0.05°C
-240 to -100°C
-100 to +40°C
+40 to +1000°C
0.1°C
0.1°C
0.05°C
0.55°C
0.2°C
0.02% RDG + 0.06°C
R -50 to +150°C
+150 to +550°C
+550 to 1768°C
0.1°C
0.1°C
0.1°C
0.95°C
0.4°C
0.02% RDG + 0.3°C
-50 to +350°C
+350 to +900°C
900°C to +1768°C
0.1°C
0.1°C
0.1°C
0.95°C
0.4°C
0.02% RDG + 0.3°C
S -50 to +150°C
+150 to +550°C
+550 to +1768°C
0.5°C
0.1°C
0.1°C
0.85°C
0.02% RDG + 0.4°C
0.02% RDG + 0.3°C
-50 to +350°C
+350 to 900°C
+900 to +1768°C
0.1°C
0.1°C
0.1°C
0.9°C
0.02% RDG + 0.4°C
0.02% RDG + 0.3°C
B +400 to +900°C
+900 to +1820°C
0.2°C
0.1°C
0.95°C
0.5°C
+400 to +850°C
+850 to +1820°C
0.2°C
0.1°C
0.95°C
0.5°C
U -200 to -100°C
-100 to +600°C
0.05°C
0.05°C
0.35°C
0.2°C
-200 to -70°C
-70 to +600°C
0.05°C
0.05°C
0.35°C
0.2°C
L -200 to -100°C
-100°C to +900°C
0.05°C
0.05°C
0.3°C
0.2°C
-200 to -70°C
-70 to +900°C
0.05°C
0.05°C
0.3°C
0.25°C
C -20 to +900°C
+900 to 2310°C
0.1°C
0.1°C
0.3°C
0.02% RDG + 0.15°C
-20 to +900°C
+900 to 2310°C
0.1°C
0.1°C
0.35°C
0.02% RDG + 0.15°C
N -240 to -190°C
-190 to -110°C
-110 to +0°C
+0 to +1300°C
0.1°C
0.1°C
0.05°C
0.05°C
0.6°C
0.25°C
0.15°C
0.02% RDG + 0.07°C
-240 to +10°C
+10 to +250°C
+250 to +1300°C
0.1°C
0.1°C
0.05°C
0.9°C
0.2°C
0.02% RDG + 0.09°C
Platine -100 to +1400°C 0.05°C 0.3°C -100 to +1400°C 0.05°C 0.35°C
Mo +0 to +1375°C 0.05°C 0.02% RDG + 0.1°C +0 to +1375°C 0.05°C 0.25°C
NiMo/NiCo -50 to +1410°C 0.05°C 0.02% RDG + 0.35°C -50 to +1410°C 0.05°C 0.02% RDG + 0.35°C

Thermocouples G, D: For specifications, refer to the instruction manual (Available on request)
Accuracy is given for reference @ 0°C.
When using the internal reference junction (except couple B) add an additional uncertainty of 0.3 °C at 0 °C.
It is possible (thermocouple B excepted) to choose by programming the cold junction localization: External at 0°C, internal (temperature compensation of instrument’s terminals) or manually entered.
Temperature coefficient: <10% of accuracy /°C

Specifications and performances of TC 6622 @23°C ±5°C

Resistance

Function Range Resolution Accuracy / 1 year Notes
IN 400 Ω 1 mΩ 0.012% RDG + 10 mΩ Automatic detection: 2, 3 or 4 wires
3600 Ω 10 mΩ 0.012% RDG + 100 mΩ Automatic detection: 2, 3 or 4 wires
OUT 400 Ω (DC current) 1 mΩ 0.012% RDG + 30 mΩ Acceptable current: 0.1 to 1 mA
3550 Ω (DC current) 10 mΩ 0.012% RDG + 300 mΩ Acceptable current: 0.1 to 1 mA

Connection in resistance and RTDs through banana plugs or 4-pin round connector
Temperature coefficient: < 7 ppm/°C beyond reference domain
Rising time in simulation: < 1 ms
R internal: < 1 Ω
Noise VLF < 1 mV (@ F< 100 Hz)

Resistive probes: Measurement and simulation

Sensor Range (Input and Output) Resolution Accuracy / 1 year (Measurement) Accuracy / 1 year (Simulation)
Pt50 (α = 3851) -220°C to +850°C 0.01°C 0.012% RDG + 0.06°C 0.012% RDG + 0.18°C
Pt100 (α = 3851) -220°C to +850°C 0.01°C 0.012% RDG + 0.05°C 0.012% RDG + 0.12°C
Pt100 (α = 3916) -200°C to +510°C 0.01°C 0.012% RDG + 0.05°C 0.012% RDG + 0.12°C
Pt100 (α = 3926) -210°C to +850°C 0.01°C 0.012% RDG + 0.05°C 0.012% RDG + 0.12°C
Pt200 (α = 3851) -220°C to +120°C 0.01°C 0.012% RDG + 0.12°C 0.012% RDG + 0.33°C
Pt500 (α = 3851) -220°C to +1200°C 0.01°C 0.012% RDG + 0.07°C 0.012% RDG + 0.18°C
Pt1000 (α = 3851) -220°C to +760°C 0.01°C 0.012% RDG + 0.05°C 0.012% RDG + 0.08°C
Ni100 (α = 618) -60°C to 180°C 0.01°C 0.012% RDG + 0.03°C 0.012% RDG + 0.08°C
Ni120 (α = 672) -40°C to +205°C 0.01°C 0.012% RDG + 0.03°C 0.012% RDG + 0.08°C
Ni1000 (α = 618) -60°C to +180°C 0.01°C 0.012% RDG + 0.03°C 0.012% RDG + 0.08°C
Cu10 (α = 427) -50°C to 150°C 0.01°C 0.012% RDG + 0.18°C 0.012% RDG + 0.1°C
Cu50 (α = 428) -50°C to +150°C 0.01°C 0.012% RDG + 0.06°C 0.012% RDG + 0.15°C

Resistive probes measurements in 2, 3 or 4 wires: automatic recognition of number of connected wires, with indication on screen
Accuracies are given for 4-wire mounted probes
Take into account particular error of temperature sensor used and implementation conditions
Measuring current: 0.65 mA
Simulation current: 0.1 mA to 1mA
Minimal current pulse duration: < 1 ms
Temperature coefficient: < 10% of accuracy /°C

Further features

Scaling in measurement and simulation modes This function allows sensors to be corrected after a calibration. Scaling is performed using up to 10 segments, in order to fit with the real calibrated value.
Calibrated sensors A database can be created to design curves for sensor s after calibration according to the corrections mentioned on a calibration report.
Data recording Data are recorded either manually on event or automatically with programmable frequency. Data is stored with date and time and can be displayed as list or curve.
Statistical functions Continuous display of average, minimum and maximum value of the signal under monitoring, as well as number of measurements.
Simple and cyclical ramp generation Ramps can be generated by setting low and high dwell, rising and falling times, stabilization and delay times (1 to 3,600 s). Delay time enables a single user to launch the ramp and go to the control panel.
Steps simulation This mode allows predefined values to be sent with programmable amplitude and frequency.
Synthesizer This mode allows predefined values to be sent with programmable frequency.

General specifications

Size 157 x 85 x 45 mm
Weight 306 g
Display 160 x 160 pixel liquid crystal graphical display with backlite
Display of result as table of values or trend curve
Power supply 4 AA batteries 1.5 V or rechargeable Ni-Mh batteries with internal charger in option
Communication ports USB
Storage capacity 10,000 data with date and time into one or several acquisition bursts

Environmental specifications

Reference range 23°C ±5°C (RH: 45 to 75% w/o condensing)
Operating reference range -10 to 50°C (RH: 20 to 80% w/o condensing)
Limit operating range -15°C to +55°C (RH: 10 to 80% w/o condensing) (70% at 55°C)
Storage temperature limits -30°C to +60°C
Maximum height 0 to 2,200 m
IP protection IP54 according to EN60529

Safety specifications

Protections
  • Electronic protection up to 250 V for ‘voltage’ wires
  • Fuse protection for ‘current’ wires
  • Protection against ‘current’ circuit breaking during inductive resistance measurements
Class In accordance with EN 61010-1
Category II, pollution 2
Rated voltage 60 V
Chocks and vibrations EN 61010-1
EMC conformity Immunity:

  • EN 61000-4-2
  • EN 61000-4-3
  • EN 61000-4-5
  • EN 61000-4-6
  • EN 61000-4-11

EN 61000-4-4
Conducted and radiated emissions:

  • EN 55022, class B
  • EN 61000-3-2
  • EN 61000-3-3

Instrument:

TC 6621                 Handheld calibrator for thermocouples  with memory

Delivered in standard with:

  • User manual
  • 4 AA batteries
  • Protection sheath
  • Carrying strap
  • Factory test report 

TC 6622                 Handheld calibrator for resistive probes with memory

Delivered in standard with:

  • User manual
  • 4 AA batteries
  • Protection sheath
  • Carrying strap
  • Factory test report

Probes for TC 6621:

ER 48145-130         Male compensated plug type T

ER 48145-140         Male compensated plug type J

ER 48145-150         Male compensated plug type S

ER 48145-160         Male compensated plug type K

ACC-A-R               Male compensated plug type LNRBEUC or D

T101                      Flexible type K sensor

T102                      Rigid K type sensor

T104                      Soft K sensor

T105                      Penetration K sensor

T106                      Surface K sensor

T703A                    Surface temp. sensor K couple

T704                      Surface K sensor + springplate

Probes for TC 6622:

ER48493-000         4-pin LEMO connector for Pt100*

S101E                   Pt100 environment sensor LEMO connector

S102E                   Pt100 immersion sensor LEMO connector

S103E                    Flexible housing sensor LEMO connector

*This accessory is necessary for RTDs with bare wires

Accessories:

AC6908                 Soft carrying case for hand-held instruments

AN6011                 Charger + batteries for hand-held instruments

ER 49519-000         USB cable mini B

Software:

DATACAL TCTM     Calibration software for TC / TM series

Supplied with USB cable mini B

Certification:

QMA11EN              COFRAC certificate of calibration

With all relevant data points where the device has been tested

Packing information:

Size                       157 x 85 x 45 mm

Weight without packing   306 g

Product Description Release Type
DATACAL

Data management software

2.10 exe
DATACAL

Instruction manual in English

2.7 pdf
DATACAL

Manuale d’istruzione

2.7 pdf
TC / TM FIRMWARE

Firmware

C.04 exe
TC / TM INSTR MANUAL

Instruction manual

NTA47230-300A7 pdf
TC / TM INSTR MANUAL GE:SP:IT

Instruction manual in German, Spanish and Italian

NTA47230-600A7 pdf
TC / TM SCPI

SCPI commands

V1.2 pdf