GreenLine 6000

picto fleche Hand-held industrial combustion gas analyzer with 6 sensors

GreenLine 6000 is a flue gas analyzer that measures up to 6 gases: O2, CO, NO, NO2, SO, CxHy with electrochemical sensors, and can calculate others like CO2. It accepts up to 10 different preloaded fuels. Further fuels can be added using GasConfig PC software.

Ask a question
Get a call

GreenLine 6000 is a flue gas analyzer that measures up to 6 gases: O2, CO, NO, NO2, SO, CxHy with electrochemical sensors, and can calculate others like CO2. It accepts up to 10 different preloaded fuels. Further fuels can be added using GasConfig PC software.

Several flue gas sampling probes are available, with different lengths and shapes, in order to match all the requirements of various applications. The sampling probe is connected to the instrument using a single or dual hose through a water trap and a suspended particle line filter. An optional heated industrial gas probe with heated hose is recommended for long-term high temperature measurements to dry the flue gas and avoid dilution of NO2 and SO2. A sintered pre‐filter is suggested for severe and heavy applications.

The gas analyzer is made of two parts communicating through cable or Bluetooth protocol: the Main Control Unit (MCU) and the Remote Control Unit (RCU). The MCU takes in charge gas analyzis while the RCU is used to display the measured data, store the analyzis in its internal memory, print out any data using the built-in impact printer and transfer data to computer.

GreenLine 6000 provides various additional functionalities available in option to enable comprehensive combustion gas analyses:

  • Air and boiler surface temperature measurements, using specific probes
  • Draft measurement
  • Automatic CO sensor internal protection from high CO level by extension of the measuring range up to 10% CO
  • Flow measurement: Measuring the flow of gas extracted from the boiler and adjusting the pump in order to obtain a constant sampling volume
  • Differential pressure measurement:  Measuring pressure at the burner and checking stack draft
  • Two 4-20 mA inputs
  • Smoke index measurement: Checking the presence of soot in gas pipes behind the heat exchanger, with a manual smoke pump, smoke index reference table and paper filter.
  • Gas velocity measurement, using a Pitot tube
  • Gas leak detection, using a leak test kit

With gas and air temperature measurements combined to gas analyzis, GreenLine 4000 will calculate the boiler efficiency, excess air and CO2 concentration.

Key features:

  • Accepts up to 6 gas sensors in different ranges, O2, CO, NO, NO2, SO, CxHy
  • Measures up to 10 different fuels in standard. Other fuels can be added using GasConfig software
  • Up to 6 sound and visual alarms if set limits of gas type and levels are exceeded.
  • DBGas 2004 software package allows the operator to easily manage all of the data & analyzis information
Measured parameter Range Resolution GreenLine 6000
O2 From 0 to 25% vol. 0.1 % EC
CO From 0 to 8000 ppm 1 ppm EC
CO From 0 to 20000 ppm 1 ppm EC
CO LOW From 0 to 500 ppm 0.1 ppm EC
CO From 0 to 10% 0.01% EC
CO2 From 0 to 99.9% 0.1% Calc
NO From 0 to 4000 ppm 1 ppm EC
NO LOW From 0 to 500 ppm 0.1 ppm EC
NOx From 0 to 4000 ppm 1 ppm Calc
NO2 From 0 to 1000 ppm 1 ppm EC
NO2 LOW From 0 to 100 ppm 0.1 ppm EC
SO2 From 0 to 4000 ppm 1 ppm EC
CxHy From 0 to 5% 0.01% Pellistore
H2S From 0 to 1000 ppm 1 ppm EC
Efficiency From 1 to 99.9% 0.1% Calc
Excess air From 1 to infinite 0.01% Calc
Delta P ±100 hPa 0.01 hPa
Tair/Tgas -10...100°C / 0...1000°C 0.1°C Pt100 / Tc K
Calc

Further features

Zero calibration Automatic calibration procedure at instrument power-on. Fresh air inlet with electrovalve and separate pneumatic circuit
Self-diagnosis Sensor efficiency test with diagnostic page
Gas level alarms Programmable from PC with GasConfig software
Sampling pump 2.2 l/min - 220 mbar with electronic flow controller
Smoke measurement Using the heated probe or the optional external manual pump
Index memory store and printout capability as standard

General specifications

Size MCU: 455 x 205 x 365 mm
RCU: 115 x 90 x 330 mm
Weight MCU: 10 kg
RCU: 0.9 kg
Power supply 110 / 230 VAC (50/60 Hz)
Battery Type: Rechargeable 7.2 Ah
Battery life: 10 hours continuous use
Communication ports RS 232
Bluetooth
Internal test memory Up to 1,000 complete analysis data points structured by Tags

Environmental specifications

Operating reference range From -5°C to +45°C (up to 50°C for short time)
Storage temperature limits From -20 to +60°C (3 months max. at temperatures exceeding the operational limits)

Instrument:

Ordering code structure: 7846 – A – B – C – D – E – F – L – M – NNNNN – P / X

Please select the required options from tables A, B… below to define the right device reference.

7846                 GreenLine 6000

Standard delivery includes:

  • Base unit with 2 sensors (O2 + CO)
  • Li-ION battery + charger module
  • Internal impact printer
  • 1,000 data memory & real time clock
  • Configuration software GasConfig
  • Data management software DBGas2004 Lite version
  • USB adapter
  • Draft and differential pressure measurement
  • Remote module with 2.5 m long shielded cable
  • Factory test report
  • Instruction manual
  • Warranty: 1 year

OR

7846WL            GreenLine 6000 Wireless (Bluetooth MCU/RCU)

 

Table A – SENSOR No. 1

1                      O2 (0 - 25%)

Table B – SENSOR No. 2

2H                    CO (0 - 8,000 ppm) H2 compensated + Dilution system

2X                     CO (0 - 20,000 ppm)

2LO                  CO (0 - 500 ppm / res. 0.1 ppm) + Dilution system

Table C – SENSOR No. 3

0                      None

4                      NO (0 - 4,000 ppm / res. 1 ppm) & NOx calculation

4LO                 NO (0 - 500 ppm / res. 0.1 ppm) & NOx calculation

Table D – SENSOR No. 4

0                      None

5                      NO2 (0 - 1,000 ppm)

5LO                  NO2 (0 - 100 ppm)

8                      CxHy (0 - 5%)

Table E – SENSOR No. 5

0                      None

6                      SO2 (0 - 4,000 ppm)

8                      CxHy (0 - 5%)

Table F – SENSOR No. 6

0                      None

7                      CO% (0 - 10%)

8                      CxHy (0 - 5%)

9                      H2S (0 - 1,000 ppm)

Table L – Sample probes

0                      None                                                                                    

1                      Ø8 / 300 mm gas + draft probe (dual hose)

2P                    Ø10 / 750 mm gas + draft probe (dual hose) removable tip         

3P                    Ø10 / 1500 mm gas + draft probe (dual hose) removable tip

2SP                  Ø10 / 750 mm heated* gas + draft probe (dual hose) removable tip

3SP                  Ø10 / 1500 mm heated* gas + draft probe (dual hose) removable tip

F10                   Sintered filter for 2, 3, S & P probes Ø10 mm 

* Heated gas probe used for smoke index measurement.

Table M – Mains adapter / Charger

1                      115 V (50/60 Hz) – USA plug

2                      230 V (50/60 Hz) – Schuko plug

3                      230 V (50/60 Hz) – UK plug

4                      230 V (50/60 Hz) – European plug

5                      100 V (50/60 Hz) – USA / Japan plug

Table N – Accessories

0                      None

1                      DBGas 2004 standard software

2                      300 mm Pitot tube

3                      750 mm Pitot tube

4                      Pt100 remote combustion air temperature probe, length: 2 m

Table P – Calibration certificate

1                      Factory certificate (gas + pressure + temperature)

Table X – Customization

INTL                 English language - manual in English - English keyboard

USA                  English language - manual in English - English keyboard

FR                    French language - manual in French - French keyboard

HOL                  Dutch language - manual in English - English keyboard

POL                  Polish language - manual in English - English keyboard

SPA                  Spanish language - manual in English - English keyboard

UK                    English language - manual in English - English keyboard

Accessories:

Probes and sensors:

BB830006          Pt100 remote air sensor + positioning cone (with 2 m cable)

BB830025          External probe for natural boiler draft (200 Pa)

BB610032          Pitot tube 300 mm

BB610033          Pitot tube 800 mm

BB610034          Pitot tube 1000 mm

EE610106          Ø8 / 300 mm tip (max. 800°C)

EE610105          Ø8 / 750 mm tip (max. 800°C)

EE610015          Ø8 / 1000 mm tip (max. 800°C)

EE610104          Ø8 / 1500 mm tip (max. 800°C)

EE610098          Ø10 / 300 mm tip (P option)

EE610102          Ø10 / 750 mm tip (P option)

EE610029          Ø10 / 1000 mm tip (P or SP option)

EE610100          Ø10 / 1500 mm tip (P option)

BB610058          Ø8 / 300 mm gas or draft probe (dual hose)

BB610066          Ø10 / 750 mm gas + draft probe (dual hose) removable tip

BB610067          Ø10 / 1500 mm gas + draft probe (dual hose) removable tip        

BB610068          Ø10 / 750 mm heated gas + draft probe (dual hose) removable tip

BB610069          Ø10 / 1500 mm heated gas + draft probe (dual hose) removable tip

F2132100          130 mm air TC K-type probe

F2137100          130 mm contact TC K-type probe

F2139000          Pipe Velcro TC K-type probe

Consult us         3 m extension for dual hose probe and Tc

Consult us         6 m extension for dual hose probe and Tc 

Test tools and miscellaneous:

EE300088          Single hose pressure probe and burner hose kit

EE300248          Kit for detecting leaks in gas networks (pump + hose + adapters)

F7828000          Manual pump for smoke index measurements + 40 filters + table

Software and connectors:

BB260166          DBGas 2004 standard software

Consumable parts:

EE340005          Printer paper roll - Minimum order of 10 pcs

EE490002          Printer ribbon - Minimum order of 3 pcs

EE650072          Line and autozero filter cartridge - Minimum order of 10 pcs

EE650073          NOx / SOx filter cartridge - Minimum order of 10 pcs

EE650074          Filter cartridge for EE650076 / EE 650082 water trap + filter - Minimum order of 10 pcs

EE650091          Quartz filter for heated head probe - Minimum order of 3 pcs

EE650099          Sintered filter for Ø10 mm tip

EE650100          Sintered filter for Ø8 mm tip

Packing information:

Size:                    Main Control Unit: 455 x 205 x 365 mm
                            Hand-Held Remote Control Unit: 115 x 90 x 330 mm

Weight:                Main Control Unit: 10 kg
                            Hand-Held Remote Control Unit: 0.9 kg

Product Description Release Type
DBGas 2004

Data management software under Windows©

4.0.18 zip
DBGas 2004

Instruction manual

MM850523-K12
GasConfig

Configuration software under Windows©

2.0.22 zip
GreenLine 6000-8000

Instruction manual

MM850371-K12 pdf