Type |
Range |
Resolution |
Accuracy (1 year) (23°C ±1°C) |
Input impedance |
Comment |
DC voltage |
200 mV 2 V 20 V 200 V 650 V |
100 μV 1 mV 10 mV 100 mV 1 V |
0.1% + 100 μV 0.1% + 1 mV 0.1% + 10 mV 0.1% + 0.1 V 0.1% + 1 V |
> 100 MΩ 1.1 MΩ 1 MΩ 1 MΩ 10 MΩ |
Optional decimal point command |
DC current |
On request 2 A |
100 mV / 1000 pts 1 mA |
0.2% + 1 mA |
Shunt 100 mV 0.1 Ω |
External Shunt 10 - 100 - 1 000 A or internal 2 - 20 - 200 mA |
AC voltage |
650 V |
1 V |
2 V |
900 kΩ |
Band-width 40 Hz to 400 Hz |
AC current |
2 A 20 A 200 A
On request
|
1 mA 10 mA 100 mA 100 mV / 1000 pts |
1% + 2 mA 1% + 20 mA 1% + 0.2 A 0.3% + 2 |
900 kΩ |
Delivered with transformer (2 - 20 ou 200 A) >For secondary of CT 1 or 5A non-included |
Process signals with scaling |
0 - 5 V 0 - 10 V 4 - 20 mA 0 - 20 mA |
Display ±2000 points For scale begin and end |
0.1% of scale + 1 |
1 MΩ 1 MΩ 50 Ω 50 Ω |
2 models, input and scale defined with order, can be modified by user (D51 model is set in factory: 4-20 mA/0-100%). an analog output is available as option. |