ICO 52 are 10,000 point programmable digital panel meters 72 x 24 mm for temperature, process voltage and process current measurements. They can be used together with temperature sensors, either resistive probes Pt or thermocouples. It is possible to change the configuration of measurement type, range, limits and scaling.
The configuration of these parameters is protected against any inopportune modifications.
- Panel meter size: 72 mm x 24 mm
- IP65 protected
- Measurement, display and scaling of:
- Temperature with Pt100, Pt1000 probes or thermocouples (7 types)
- Process voltage and current (0-10 V, 4-20 mA)
- Resistance
- 9 different power supplies available
Process electrical measurement indicator
Type |
Range (1) |
Resolution |
Accuracy (1 year) (23°C ±1°C) |
Comments |
DC voltage |
50 mV 500 mV 5 V 50 V |
10 µV 100 μV 1 mV 10 mV |
0.1% + 10 μV 0.1% + 100 μV 0.1% + 1 mV 0.1% + 10 mV |
Input impedance > 1000 MΩ > 1000 MΩ 1 MΩ 1 MΩ |
DC current |
4-20 mA or 0-50 mA |
10 µA |
0.2% + 10 µA |
10 Ω shunt |
AC current |
1 A (2) 5 A (2) |
1 mA 1 mA |
|
Measuring current 200 µA |
Temperature coefficient: ≤ 10% accuracy/°C
Automatic ranging
(1) Extended measurement range: -40 to +110% range.
Temperature indicator
Type |
Range (1) |
Resolution |
Accuracy over 1 year (23°C ±1°C) (2) |
Pt100 |
-200 to +500°C -220 to +1200°C |
0.1°C 1°C |
0.1% RDG + 0.2°C 0.1% RDG + 2°C |
Pt1000 |
-200 to +500°C |
0.1°C |
0.1% RDG + 0.2°C |
Tc K |
-100 to +1000°C +1000 to 1372°C |
0.1°C 1°C |
0.1% RDG + 0.4°C 0.1% RDG + 1°C |
Tc T |
-100 to +400°C |
0.1°C |
0.1% RDG + 0.4°C |
Tc Platine |
-100 to +1400°C |
0.1°C |
0.1% RDG + 0.4°C |
Tc J |
-120 to +999°C |
0.1°C |
0.1% RDG + 0.4°C |
Tc L |
-40 to +900°C |
0.1°C |
0.1% RDG + 0.4°C |
Tc R |
-50 to +120°C +120 to 1768°C |
1°C 1°C |
0.1% RDG + 2°C 0.1% RDG + 1°C |
Tc S |
-50 to +450°C +450 to 1768°C |
1°C 1°C |
0.1% RDG + 2°C 0.1% RDG + 1°C |
Temperature coefficient: ≤ 10% accuracy/°C
Display in °C or °F.
(1) For a defined type of sensor, the commutation of range and the change of resolution can be automatic.
Measurement range extendable.
(2) Accuracy given without reference junction.
The typical error brought by the reference junction (AN8002) is 0.3°C.
Further features
Measurement hold |
Short-circuit on the rear panel |
Scaling |
The value is to be entered using keyboard, while the value of measurement corresponding can be is introduced of the same way, is measured by the instrument. This possibility is very interesting for error corrections on temperature ranges. |
Sampling rate |
2.5 samples / s |
General specifications
Size |
99 x 36 x 120 mm |
Weight |
300 g |
Display |
From -1999 to 9999 with 14 mm red, yellow or green indication LED |
Supply |
230 V at ±10% (50/60 Hz) Other power supplies available in option |
Environmental specifications
Reference range |
23°C ±1°C (20 to 75 % w/o condensing) |
Operating reference range |
0 to 50°C (RH: 20 to 75 % w/o condensing) |
Limit operating range |
-10°C to +50°C (RH: 10 to 80 % w/o condensing) |
Indice de protection |
IP40 in ABS case IP65 in watertight metallic case |
Instrument:
Ordering reference code: ICO52-Isolated power supply-Display-Enclosure
Isolated power supply:
48 V~ 9
230 V~ 8
115 V~ 7
24 V~ 6
36 to 72 V- 5
18 to 36 V- 4
9 to 18 V- 3
5 V- 2
Display:
Red R
Green V
Enclosure:
Plastic P
Metal M
Accessories:
AN8002 Cold Junction Compensation Module
AN8003 Shunt 1 A - 100 mV
AN8004 Shunt 5 A - 100 mV
AN8005 Shunt 4-20 mA (50 ohms)
AN5844 Current transformer 2/1 A
AN5845 Current transformer 20/1 A
AN5846 Current transformer 200/1 A
ER42062-001 RC protection circuit
AMD002-000: Current transformer 20/5A
AMD003-000: Current transformer 200/5 A
VISULOG |
Instruction manual
|
|
pdf
|
VISULOG WEB |
For remote access to VISULOG software
|
1.5B |
exe
|