ICO 61
Programmable digital panel meter for electrical signals
ICO 61 is a 10,000 programmable digital panel meter 96 x 24 mm for electrical signals.
ICO 61 is a 10,000 programmable digital panel meter 96 x 24 mm for electrical signals.
ICO 61 is a 10,000 programmable digital panel meter 96 x 24 mm for electrical measurements. It is possible to change the configuration of measurement type, range, limits and scaling. The configuration of these parameters is protected against any inopportune modifications.
Type | Range | Resolution | Accuracy (1 year) (23°C ±1°C) |
Input impedance |
DC voltage | 500 mV 5 V 50 V 400 V |
100 μV 1 mV 10 mV 100 mV |
0.1% + 100 μV 0.1% + 1 mV 0.1% + 10 mV 0.1% + 0.1 V |
> 100 MΩ 1.1 MΩ 1 MΩ 1 MΩ |
AC voltage | 500 mV 5 V 50 V 400 V |
100 μV 1 mV 10 mV 100 mV |
0.7% + 1 mV 0.5% + 6 mV 0.5% + 60 mV 0.5% + 0.6 V |
1 MΩ 1 MΩ 1 MΩ 1 MΩ |
DC current | 4-20 mA or 0-50 mA 1 A (2) 5 A (2) |
10 µA 1 mA 1 mA |
0.2% + 10 µA | shunt 10 Ω included external shunt 100 mV external shunt 100 mV |
AC current | 1 A (2) 5 A (2) |
1 mA 1 mA |
external shunt 100 mV external shunt 100 mV |
|
Frequency (3) | 5 to 500 Hz 5 to 5000 Hz |
0.1 Hz 1 Hz |
0.1% + 0.1 Hz 0.1% + 1 Hz |
10 MΩ 10 MΩ |
Temperature coefficient: ≤ 10% accuracy/°C
Continuous mode: -40 to 100 % range
Alternate mode: 4 to 100 % range from 25 to 500 Hz without D.C component (possible measurements up to 10 kHz on some ranges with less precision)
(1) Extended measurement range: -40 to +110% range
(2) With external shunts
(3) Two levels of input voltage: 50 mV~ to 50 V~ and 5 V~ to 250 V~
Frequency:
- 70 V peak on low entry level (50 mV to 50 V)
- 570 V peak on entry high level (5 V to 250 V)
Ranging | Automatic |
Measurement hold | Short-circuit on the rear panel |
Scaling | The value is to be entered using keyboard, while the value of measurement corresponding can be is introduced of the same way, is measured by the instrument. This possibility is very interesting for error corrections on temperature ranges. |
Sampling rate | 2.5 samples / s |
Size | 99 x 36 x 120 mm |
Weight | 300 g |
Display | From -1999 to 9999, with 14 mm red, yellow or green indication LED |
Supply | 230 V at ±10% (50/60 Hz) Other power supplies available in option |
Reference range | 23°C ±1°C (20 to 75 % w/o condensing) |
Operating reference range | 0 to 50°C (RH: 20 to 75 % w/o condensing) |
Limit operating range | -10°C to +50°C (RH: 10 to 80 % w/o condensing) |
Indice de protection | IP40 in ABS case IP65 in watertight metallic case |
Protections | - Protection of unit programming by access code. - Protection of the threshold values by safety function programming |
Ordering reference code: ICO51-Power supply-Display-Enclosure
Power supply:
48 V~ 9
230 V~ 8
115 V~ 7
24 V~ 6
36 to 72 V- 5
18 to 36 V- 4
9 to 18 V- 3
5 V- 2
Display:
Red R
Green V
AN8002 Cold Junction Compensation Module
AN8003 Shunt 1 A - 100 mV
AN8004 Shunt 5 A - 100 mV
AN8005 Shunt 4-20 mA (50 ohms)
AN5844 Current transformer 2/1 A
AN5845 Current transformer 20/1 A
AN5846 Current transformer 200/1 A
ER42062-001 RC protection circuit
AOIP offers a complete range of services: technical assistance and advice, calibration and verification by the SOFIA laboratory, equipment repair, maintenance, assistance and metrological monitoring, equipment rental, training, installation and commissioning of your equipment. More informations